Buccutite™ Rapid PE Antibody Labeling Kit *Production Scale Optimized for Labeling 1 mg Antibody Per Reaction*
Example protocol
AT A GLANCE
1.0 mg Antibody (MW ~150 kDa)
Antibody concentration: 2.0 mg/mL
Antibody volume: 500 µL
PREPARATION OF WORKING SOLUTION
Before opening the vials, warm all components and briefly centrifuge. Immediately prepare necessary solutions before starting conjugation. This protocol is a recommendation.
Prepare a 500 µL antibody solution in PBS with a concentration of 2 mg/mL.
Note: The protein should be dissolved in 1X phosphate-buffered saline (PBS), pH 7.2 - 7.4. If the protein is dissolved in buffers containing primary amines, like Tris and/or glycine, it must be dialyzed against 1X PBS, pH 7.2 - 7.4, or use Amicon Ultra0.5, Ultracel-10 Membrane, 10 kDa (Cat No. UFC501008 from Millipore) to remove free amines or ammonium salts (such as ammonium sulfate and ammonium acetate) that are widely used for protein precipitation.
Note: Impure antibodies or antibodies stabilized with bovine serum albumin (BSA) or gelatin will not be labeled well.
Warm up a vial of Buccutite™ MTA (Component B) to room temperature.
Add 5 µL of DMSO (not provided) to the vial of Buccutite™ MTA (Component B), and mix well by pipetting.
SAMPLE EXPERIMENTAL PROTOCOL
Add 25 µL of Reaction Buffer (Component C) to the antibody solution.
Transfer 5 µL of the reconstituted Buccutite™ MTA DMSO solution into the vial of antibody solution, and mix well by pipetting.
Rotate the reaction mixture at room temperature for 1 hour, then purify using a desalting column.
Invert the provided spin column (Component D) several times to re-suspend the settled gel and remove any bubbles.
Snap off the tip and place the column in a washing tube (2 mL, not provided). Remove the cap to allow the excess packing buffer to drain by gravity to the top of the gel bed.
Note: If the column does not begin to flow, push the cap back into the column and remove it again to start the flow. Discard the drained buffer, and then place the column back into the Washing Tube.
Centrifuge at 1000 x g for 2 minutes in a swinging bucket centrifuge to remove the packing buffer. Then discard the buffer. Refer to the 'Centrifugation Notes' section below for instructions.
Apply 1-2 mL 1X PBS (pH 7.2-7.4) to the column. After each application of PBS, let the buffer drain out by gravity, or centrifuge the column for 2 minutes to remove the buffer. Discard the buffer from the collection tube. Repeat this process for 3-4 times.
Centrifuge at 1000 x g for 2 minutes in a swinging bucket centrifuge to remove the packing buffer. Then discard the buffer. Refer to the 'Centrifugation Notes' section below for instructions.
Place the column into a clean collecting tube (1.5 mL, not provided). Then, take the antibody-Buccutite™ MTA solution from step 3 of the "Run Antibody-Buccutite™ MTA Reaction" section and load it carefully and directly into the center of the column.
After loading the sample, add 40 μL of 1X PBS (pH 7.2-7.4), centrifuge the column for 2 minutes at 1,000 x g, and collect the solution that contains the desired antibody-Buccutite™ MTA solution.
Warm up a vial of Buccutite™ FOL-Activated PE (Component A) to room temperature.
Note: Each vial of Buccutite™ FOL-Activated PE contains an optimized amount of dye to label 1 mg of IgG (MW ~150 kDa) at 2 mg/mL in PBS, the kit can also be used to label other proteins (>10 kDa).
Make a Buccutite™ FOL-Activated PE solution by adding 250 µL of ddH2O into the vial of Buccutite™ FOL-Activated PE (Component A), and mix well by pipetting or vortexing.
Add the purified Antibody-Buccutite™ MTA solution directly into the vial of Buccutite™ FOL-Activated PE solution. Rotate the mixture for 1-2 hours at room temperature.
The antibody-PE conjugate is now ready for immediate use or can be stored at 4°C.
For optimal performance, it is recommended to purify the antibody-PE conjugate using size exclusion chromatography (SEC). The following SEC columns are suitable for this purpose: Superdex 200 Increase 100/300 GL (Cytiva) and ENrich™ SEC 650 10 x 300 Column (Bio-Rad).
Spectrum
Product family
Name | Excitation (nm) | Emission (nm) | Extinction coefficient (cm -1 M -1) | Correction Factor (280 nm) |
Buccutite™ Rapid APC Antibody Labeling Kit *Microscale Optimized for Labeling 100 ug Antibody Per Reaction* | 651 | 660 | 730000 | 0.195 |
Buccutite™ Rapid APC Antibody Labeling Kit *Microscale Optimized for Labeling 25 ug Antibody Per Reaction* | 651 | 660 | 730000 | 0.195 |
Buccutite™ Rapid PerCP Antibody Labeling Kit *Microscale Optimized for Labeling 100 ug Antibody Per Reaction* | 477 | 678 | 406000 | 0.22 |
Buccutite™ Rapid PerCP Antibody Labeling Kit *Microscale Optimized for Labeling 25 ug Antibody Per Reaction* | 477 | 678 | 406000 | 0.22 |
Buccutite™ Rapid PerCP Antibody Labeling Kit *Production Scale Optimized for Labeling 1 mg Antibody Per Reaction* | 477 | 678 | 406000 | 0.22 |
References
Authors: Lian, Cheng and Young, Dan and Randall, Richard E and Samuel, Ifor D W
Journal: Biosensors (2022)
Authors: Renner, Patrick and Crone, Michael and Kornas, Matthew and Pioli, KimAnh T and Pioli, Peter D
Journal: Antibody therapeutics (2022): 151-163
Authors: Muench, Marcus O and Nosworthy, Christopher
Journal: BMC research notes (2022): 358
Authors: Chitadze, Guranda and Lettau, Marcus and Peters, Christian and Luecke, Stefanie and Flüh, Charlotte and Quabius, Elgar Susanne and Synowitz, Michael and Held-Feindt, Janka and Kabelitz, Dieter
Journal: Cytometry. Part B, Clinical cytometry (2022): 228-238
Authors: Ugamraj, Harshad S and Dang, Kevin and Ouisse, Laure-Hélène and Buelow, Benjamin and Chini, Eduardo N and Castello, Giulia and Allison, James and Clarke, Starlynn C and Davison, Laura M and Buelow, Roland and Deng, Rong and Iyer, Suhasini and Schellenberger, Ute and Manika, Sankar N and Bijpuria, Shipra and Musnier, Astrid and Poupon, Anne and Cuturi, Maria Cristina and van Schooten, Wim and Dalvi, Pranjali
Journal: mAbs (2022): 2095949