Amplite® Colorimetric Endotoxin Detection Kit

Lipopolysaccharide (LPS), also known as endotoxin, is the major component of the outer membranes of Gram-negative bacteria. LPS is a potent stimulator of the vertebrate innate immune system and can cause fever, septic shock and eventually death. It is also recognized as a biomarker for the detection of bacterial pathogen invasion, and responsible for the development of inflammatory response and endotoxic shock in extreme cases. Detection of LPS in biological materials, such as protein, peptide or antibody sample, is a critical task in biomanufacturing and bioprocessing. Amplite®™ Colorimetric Endotoxin Detection Kit uses Endotoxin Yellow™, a sensitive chromogenic substrate. Endotoxin Yellow™ can be hydrolyzed in the presence of endotoxins and the Limulus Amebocyte Lysate (LAL), an extract of blood cells from a horseshoe crab, to generate an intense yellow colored product. The endotoxin activity is proportional to the absorbance of the yellow product resulted from the hydrolysis of Endotoxin Yellow™. Amplite®™ Colorimetric Endotoxin Detection Kit can detect a broad range of endotoxin from 1 EU/ml to 0.002 EU/ml.

Example protocol

AT A GLANCE

Protocol summary
  1. Prepare Endotoxin Yellow™ working solution
  2. Add E.coli Endotoxin Standards and test samples (25 µL)
  3. Add Limulus Amebocyte Lystate solution (25 µL)
  4. Incubate at 37 °C for 30 minutes
  5. Add Endotoxin Yellow™ working solution (50 µL)
  6. Read optical density at 498 nm within 10 minutes
Important

Thaw all the kit components at room temperature before starting the experiment.

All Materials used in the experiment should be endotoxin-free, such as: disposable tubes or 1.5 mL microcentrifuge tubes, disposable pipette tips, and disposable 96-well microplates or plate strips. The cleanliness of all labware is required to accurately detect levels of endotoxin in a given sample.

PREPARATION OF STOCK SOLUTIONS

Unless otherwise noted, all unused stock solutions should be divided into single-use aliquots and stored at -20 °C after preparation. Avoid repeated freeze-thaw cycles

100X Endotoxin Yellow™ stock solution

Add 50 µL of DMSO into the vial of Endotoxin Yellow™  (Component A) to make 100X Endotoxin Yellow™ stock solution.
Note        Keep from light.

Limulus Amebocyte Lysate (LAL) Stock Solution

Add 500 µL Endotoxin-Free Water (Component B) to the vial of Limulus Amebocyte Lysate (Component C) to make 5X  Limulus Amebocyte Lysate (LAL) stock solution.

PREPARATION OF STANDARD SOLUTIONS

For convenience, use the Serial Dilution Planner:
https://www.aatbio.com/tools/serial-dilution/60007

E.coli Endotoxin Standard solution
Add 40 µL of 100 EU/mL E.coli Endotoxin Standard solution to 360 µL of Endotoxin-Free Water (Component B) to generate 10 EU/mL E.coli Endotoxin Standard solution (ES1). Then take 10 EU/mL E.coli endotoxin standard solution (ES1) and perform 1:2 serial dilutions in Endotoxin-Free Water (Component B) to get serially diluted E.coli Endotoxin Standards (ES2 - ES7).

PREPARATION OF WORKING SOLUTION

Endotoxin Yellow™ working solution

Add 50 µL of Endotoxin Yellow™ stock solution into 5 mL of Endotoxin-Free Water (Component B) to make a total volume of 5.05 mL Endotoxin Yellow™ working solution.
Note        Prepare the amount of endotoxin substrate working solution as needed. Keep the working solution from light.

Limulus Amebocyte Lysate (LAL) working solution

Add 500 µL of  Limulus Amebocyte Lysate (LAL) Stock Solution into 2 mL of Endotoxin-Free Water (Component B) to make a total volume of 2.5 mL Limulus Amebocyte Lysate (LAL) working solution.
Note        Prepare the amount of LAL working solution as needed and before use. Using the Endotoxin-Free bottle or tube.

SAMPLE EXPERIMENTAL PROTOCOL

Table 1. Layout of E.coli endotoxin standards and test samples in a clear bottom 96-well microplate. ES=E.coli endotoxin standards (ES1-ES7, 1.00 to 0.001 EU/mL); BL=Blank Control; TS=Test Samples

BLBLTSTS
ES1ES1......
ES2ES2......
ES3ES3  
ES4ES4  
ES5ES5  
ES6ES6  
ES7ES7  

Table 2. Reagent composition for each well.

WellVolumeReagent
ES1-ES725 µLSerial dilutions (1 to 0.001 EU/mL)
BL25 µLEndotoxin-Free Water
TS25 µLTest Samples
  1. Prepare E.coli Endotoxin Standards (ES), blank controls (BL), and test samples (TS) according to the layout provided in Tables 1 and 2. For a 384-well plate, use 12.5 µL of reagent per well instead of 25 µL.
  2. Add 25 µL of Limulus Amebocyte Lysate solution to each well of E.coli Endotoxin Standard, blank control and test samples.
  3. Mix well and incubate for 30 minutes at 37 °C.
  4. Add 50 µL of Endotoxin Yellow™ working solution to each well of E.coli Endotoxin Standard, blank control, and test samples to make the total assay volume 100 µL/well. For a 384-well plate, add 25 µL of Endotoxin Yellow™ solution into each well instead, for a total volume of 50 µL/well.
  5. Monitor the optical density at 498 nm..

    Note        For best results, read between 2 to 10 minutes after adding the working solution.

    Note        25 µL of 25% acetic acid can be added to stop the reaction.

Citations

View all 35 citations: Citation Explorer
Exosomal Tenascin-C Primes Macrophage Pyroptosis Amplifying Aberrant Inflammation during Sepsis-Induced Acute Lung Injury
Authors: Gong, Ting and Zhang, Xuedi and Liu, Xiaolei and Ye, Yinfeng and Tian, Zhiyuan and Yin, Shuang and Zhang, Min and Tang, Jing and Liu, Youtan
Journal: Translational Research (2024)
Macrophage-derived exosomal aminopeptidase N aggravates sepsis-induced acute lung injury by regulating necroptosis of lung epithelial cell
Authors: Gong, Ting and Zhang, Xuedi and Peng, Zhiyong and Ye, Yinfeng and Liu, Ruimeng and Yang, Yinggui and Chen, Zhugui and Zhang, Zhihao and Hu, Hongfei and Yin, Shuang and others,
Journal: Communications Biology (2022): 1--17
A Validation Study of the Limulus Amebocyte Lysate Test as an End-Product Endotoxin Test for Polyvalent Horse Snake Antivenom
Authors: Sheraba, N. S., Diab, M. R., Yassin, A. S., Amin, M. A., Zedan, H. H.
Journal: PDA J Pharm Sci Technol (2019): 562-571
Utility Of An Automatic Limulus Amebocyte Lysate Kinetic Turbidimetric Test For Endotoxin Screening Of Dialysate Samples
Authors: Uchida, T., Kaku, Y., Hayasaka, H., Kofuji, M., Momose, N., Miyazawa, H., Ueda, Y., Ito, K., Ookawara, S., Morishita, Y.
Journal: Med Devices (Auckl) (2019): 429-433
Evaluation of a portable test system for assessing endotoxin activity in raw milk
Authors: Suzuki, Y., Suzuki, K., Shimamori, T., Tsuchiya, M., Niehaus, A., Lakritz, J.
Journal: J Vet Med Sci (2016): 49-53
Page updated on December 17, 2024

Ordering information

Price
Unit size
Catalog Number60007
Quantity
Add to cart

Additional ordering information

Telephone1-800-990-8053
Fax1-800-609-2943
Emailsales@aatbio.com
InternationalSee distributors
Bulk requestInquire
Custom sizeInquire
Technical SupportContact us
Purchase orderSend to sales@aatbio.com
ShippingStandard overnight for United States, inquire for international
Request quotation

Storage, safety and handling

H-phraseH303, H313, H333
Hazard symbolXN
Intended useResearch Use Only (RUO)
R-phraseR20, R21, R22
UNSPSC12171501

Platform

Absorbance microplate reader

Absorbance498 nm
Recommended plateClear bottom

Components

E.coli Endotoxin dose response was measured in an all-clear 96-well plate using a with a Spectrum Max microplate reader (Molecular Devices) at OD 498nm. As low as 0.001 EU/mL of E.coli Endotoxin can be detected with 10 minutes incubation (n=3).
E.coli Endotoxin dose response was measured in an all-clear 96-well plate using a with a Spectrum Max microplate reader (Molecular Devices) at OD 498nm. As low as 0.001 EU/mL of E.coli Endotoxin can be detected with 10 minutes incubation (n=3).
E.coli Endotoxin dose response was measured in an all-clear 96-well plate using a with a Spectrum Max microplate reader (Molecular Devices) at OD 498nm. As low as 0.001 EU/mL of E.coli Endotoxin can be detected with 10 minutes incubation (n=3).